For testing the residual seal force of flanged caps, Zwick recommends a device with inserts and compression dies of various sizes. Measuring the pre-load force of the rubber stopper on an aluminium flanged cap provides an indication of the security of the vial closure. Modular design allows the various inserts to be changed quickly and easily; a protective screen is necessary for safety purposes, as glass breakage may occur during this test.